International Light's new IL 900 System

System Features & Benefits

* NIST-Traceable Spectroradiometer from 220 to 1050 nm at 1 nm resolution
* Spectral sensitivity better than:
o Visible: 5 nanoWatt/cm2/nm
o UV/NIR: 0.1 microWatt/cm2/nm
* Stray light rejection: better than 5 x 10-4 of total integrated power per nm
* Intrinsic Repeatability: 0.5%
* Spectral accuracy: 0.5 nm
* Radiometric accuracy: ± 3 % VIS, ± 5% UV/NIR
* Cosine Receptor accuracy: ± 3% from 0 - 90°, all wavelengths
* Rugged, shock-mounted construction in dustproof die cast aluminum housing
* Temperature range: 15 - 40°C
* Standard stainless steel armored 2m coupling fiber
* 2048 element CCD with proprietary order-sorting filtration
* 16-bit conversion resolves low-level spectra on large backgrounds
* Exposure time adjustable from 1 mS to 5 S
* Data collection is software controlled or externally triggered
* USB 2.0 interface, powered by USB connection
* Requires: Windows 98, 2000, XP, Pentium II 300 MHz or better
* Available in three options

Spectrilight™ Control Software

The SpectrILight software package puts complete control of data taking in the user’s hands. Wavelength range and resolution, exposure time, scan averaging and smoothing, batch and pre-scheduled data collection are all handled. The % Transmission/Reflectance and Irradiance displays can be exported to ASCII text, Excel®, BMP or SPC formats, complemented by a suite of built-in report generating functions. On-line reflectance color analysis is also provided, including Chromaticity and Color Difference.

Spectrilight Tools™

In addition to the on-line analysis available in SpectrILight, SpectrILight Tools is a powerful Excel® app that accepts 1 nm resolution input files and calculates all standard spectral parameters such as lux, scotopic intensity, UVA,B,C,VIS, PAR including Einsteins, Chromaticity, Dominant and Complementary wavelength, Color Purity, CCT, CRI, Solar Simulator and more. User-selectable bands can be defined interactively, and userdefined weighting functions can also be installed.

For more information please contact Mr. Bart Hengsdijk email: bhengsdijk@laser2000.nl

 